The study
Between 15 May and 30 November 2025, six NHS microbiology laboratories ran Halo AI on positive blood cultures from their neonatal units. It was a prospective, multicentre diagnostic accuracy study. Halo AI ran as a laboratory prototype and was compared with ISO 20776-1 broth microdilution, read against EUCAST breakpoints. Clinicians did not see its results.
Rose EG, Mensah DK, Raman P, Fenwick A, Okoro A. Single-cell phenotypic antimicrobial susceptibility testing of neonatal positive blood cultures in under two hours: a prospective, multicentre diagnostic accuracy study. Preprint, January 2026.
Headline figures
1,214 cultures 4,842 organism–antibiotic pairs
-
96.8%
Categorical agreement 95% CI 96.3–97.3
-
0.7%
False-susceptible results 8 of 1,11695% CI 0.4–1.4 Halo AI called a resistant organism susceptible. The paper uses the testing standard’s term for this, very major errors. The whole confidence interval is below the accepted limit of 1.5%.
-
1.6%
False-resistant results 60 of 3,726 Halo AI called a susceptible organism resistant.
Agreement by organism
| Organism | Pairs | Agreement | False susceptible | False resistant |
|---|---|---|---|---|
| Coagulase-negative staphylococci | 1,648 | 96.2% | 4 of 721 | 22 of 927 |
| S. aureus | 672 | 97.3% | 2 of 152 | 7 of 520 |
| Group B streptococcus | 588 | 98.0% | none resistant | 4 of 588 |
| E. coli | 1,270 | 97.3% | 2 of 200 | 14 of 1,070 |
| Klebsiella | 448 | 95.8% | 0 of 35 | 8 of 413 |
| Enterobacter | 216 | 95.4% | 0 of 8 | 5 of 208 |
| All organisms | 4,842 | 96.8% | 8 of 1,116 | 60 of 3,726 |
Time to result
Each bar is the interquartile range and the line across it is the median, in minutes from loading the cartridge. The red edge marks 2 hours.
| Organism | Median (IQR) | Reported within 2 h |
|---|---|---|
| Coagulase-negative staphylococci | 102 min (89–118) | 83.3% |
| S. aureus | 74 min (65–86) | 98.2% |
| Group B streptococcus | 86 min (76–97) | 95.9% |
| E. coli | 52 min (46–61) | 100% |
| Klebsiella | 58 min (51–67) | 100% |
| Enterobacter | 63 min (55–72) | 100% |
| All organisms | 88 min (63–104) | 93.4% |
Compared with routine testing
From the moment a culture flagged positive, routine testing at the study sites took a median of 46 hours to give a result. Halo AI took a median of 88 minutes from loading the cartridge.
Routine testing 46hours
Halo AI 88minutes
Routine testing is counted from the moment the culture flagged positive, and Halo AI from loading the cartridge. The bars are to scale.
Limitations
- We did not report results to clinicians, so the study can’t show any effect on treatment or outcomes. That will take a randomised evaluation, which is the next study we plan.
- Coagulase-negative staphylococci are common contaminants, so some of the cultures we included will not have been real infections.
- Few Klebsiella and Enterobacter isolates were resistant, so the error estimates for those two organisms are imprecise.
- Every site was in England, where resistance is lower than in the countries where neonatal sepsis kills the most babies.
- The study ran on a laboratory prototype. The benchtop Halo Reader has not been built yet, and when it is we will have to confirm that it performs as the prototype did.